Reagent icons, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 172

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 172
Reagent icons
Indicates the state of the reagent system. This icon is displayed at the
bottom of the Reagent window.
Sunny icon:
The reagent calibration or manual entry of the titre value
has been performed on all the reagents present in the
system.
Cloudy icon:
The reagent calibration of one of the reagents present in
the system should be performed within 12* or 24 hours.
The expiry date of one of the reagents in the system will
expire in less than one week.
* When the Periodicity is set to 1 day, this icon will appear
to indicate that a calibration must be performed within
12 hours.
Stormy icon:
The reagent calibration or the expiry date has elapsed for
one of the reagents present in the system.
If acceptance limits have been set for the reagent: at least
one calibration result lies outside the programmed
acceptance limits.
At least one of the reagents present in the system has not
been installed.
Question mark:
The reagent system has not been programmed correctly.
Enter supervisor and check the reagent parameters in the
Method parameters menu. For a reagent calibration,
make sure that the electrode(s) used for the calibration
are the same as those used in the method.
Calculator icon:
Reprocessing mode (Working mode = Reprocessing) is
set on the instrument. This icon is displayed irrespective
of the status of the reagent system used.
Refer to "Working mode", page 226
Press 1 in the Main window, the instrument will indicate the
possible errors and prompt you to correct them.