Replace burette, Replace electrodes, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 177

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 177
Replace burette
Before replacing, rinse the burette to remove the reagent.
1.
Enter the Burette functions menu.
2.
Select the burette to be replaced at the Address line,
3.
Press 5 Replace burette. The piston is moved to its lowest posi-
tion (filling operation).
4.
The data of the burette to be replaced are displayed. Check and
press 1.
5.
Enter the new burette volume and serial number as for a burette
installation. Press 1 to confirm.
6.
Remove the burette and press
✓.
7.
Install the new burette at the displayed address and press
✓,
.
Replace
electrodes
Use this procedure to replace an electrode with another one of the
same type and ID.
Proceed as follows:
1.
Select the method/sequence using the electrode.
2.
Press 2 in the Electrode window.
3.
Press 3 Replace electrode.
4.
Select the electrode to be replaced. The ID list contains all the
electrodes connected.
5.
Disconnect electrode and press
✓ to confirm.
6.
Connect the new electrode at the address indicated.
7.
Enter the serial number and confirm.