Replace burette

Before replacing, rinse the burette to remove the reagent.

1.

Enter the Burette functions menu.

2.

Select the burette to be replaced at the Address line,

see

"Address", page 51

.

3.

Press 5 Replace burette. The piston is moved to its lowest posi-
tion (filling operation).

4.

The data of the burette to be replaced are displayed. Check and
press 1.

5.

Enter the new burette volume and serial number as for a burette
installation. Press 1 to confirm.

6.

Remove the burette and press

✓.

7.

Install the new burette at the displayed address and press

,

see

"Install burette", page 125

.

Replace