Electrode
connection

Proceed as follows to connect/install electrodes and temperature
sensors:

1.

In the Electrode window, press 2 then 1 Connect electrode.

2.

Enter serial number

3.

Install electrodes on the Titration Manager (TIM), ABU52 or
sample changer (SAC80, SAC90, SAC850 or SAC950).

4.

Connect electrodes to the rear panel socket of the Titration
Manager (TIM) or ABU52 (ABU1 or ABU2),

see "ABU1/ABU2",

page 43

. See figure and table below. For example: pHC2001 to

address TIM/E1.

Refer to "Address", page 51

.

5.

Press 1 to confirm.

Figure 17: Electrode sockets

Table 3: Connecting electrodes

If the current method in use requires an electrode different to
the one already connected, the Titration Manager will prompt
you to disconnect the electrode before connecting the new
one,

see "Disconnect electrodes", page 88

.

Refer to "Electrode connection - Important", page 103

.

P

t

-P

t

R

ef

E1

E2

T

emp

GND

Socket

Electrode

REF

Single reference

TEMP

Temperature

GND

Single metal for cell
grounding only

Pt-Pt

Double metal

E1/E2

Others