Electrode connection, Refer to "electrode connection, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 102

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 102
Electrode
connection
Proceed as follows to connect/install electrodes and temperature
sensors:
1.
In the Electrode window, press 2 then 1 Connect electrode.
2.
Enter serial number
3.
Install electrodes on the Titration Manager (TIM), ABU52 or
sample changer (SAC80, SAC90, SAC850 or SAC950).
4.
Connect electrodes to the rear panel socket of the Titration
Manager (TIM) or ABU52 (ABU1 or ABU2),
. See figure and table below. For example: pHC2001 to
address TIM/E1.
.
5.
Press 1 to confirm.
Figure 17: Electrode sockets
Table 3: Connecting electrodes
If the current method in use requires an electrode different to
the one already connected, the Titration Manager will prompt
you to disconnect the electrode before connecting the new
one,
see "Disconnect electrodes", page 88
.
Refer to "Electrode connection - Important", page 103
P
t
-P
t
R
ef
E1
E2
T
emp
GND
Socket
Electrode
REF
Single reference
TEMP
Temperature
GND
Single metal for cell
grounding only
Pt-Pt
Double metal
E1/E2
Others