Function
Select the electrode function relative to the electrode in use.
The possible electrode functions are:
•
pH,
•
mV (i = 0),
•
mV (i > 0),
•
T°C,
•
Reference,
•
Ground.
Refer to "Electrode type", page 106
Fuses
For continued protection replace the fuse with one of a high
interrupting capacity, same type and rating:
2 x fuses, slow blow, 1.0 A (5 x 20 mm), part no. 450-020.
To replace the fuses:
Figure 21: Fuse replacement
hook
3. Remove the fuse holder
4. Replace the used fuses with
ones of the same type and rating
1. Switch off the instrument
5. Put the cap back in place
2. Disconnect line cord