Stability, Standard amount, Standard id – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 211: Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 211
Stability
The stability criterion is used to control the stability of the electrode
signal. Selecting a low value will bring about, accurate but long
measurements.
Enter a stability which is close to the default value (50 mpH/min or
3.0 mV/min). A zero stability criteria will have no effect on the stability
test performed, the reading will be taken into account when the
Acceptance time is exceeded.
Enter in:
Edit method > Parameters menu
Edit electrode > Calibration parameters menu
Range available:
0 to 99 mpH/min or 0 to 99.9 mV/min
Standard
amount
Amount of standard used to calibrate the titrant.
Enter in:
Edit reagent > Standard
Range available:
0.001 to 100000 (unit = Amount unit)
Standard ID
Name assigned to the standard used to calibrate the titrant (max. 20
alphanumeric characters).
Enter in:
Edit reagent > Standard menu