Stability

The stability criterion is used to control the stability of the electrode
signal. Selecting a low value will bring about, accurate but long
measurements.

Enter a stability which is close to the default value (50 mpH/min or
3.0 mV/min). A zero stability criteria will have no effect on the stability
test performed, the reading will be taken into account when the
Acceptance time is exceeded.

Enter in:

Edit method > Parameters menu
Edit electrode > Calibration parameters menu

Range available:

0 to 99 mpH/min or 0 to 99.9 mV/min

Standard
amount

Amount of standard used to calibrate the titrant.

Enter in:

Edit reagent > Standard

Range available:

0.001 to 100000 (unit = Amount unit)

Standard ID

Name assigned to the standard used to calibrate the titrant (max. 20
alphanumeric characters).

Enter in:

Edit reagent > Standard menu