Ip reject, Ip>1 (result indicator), Iso ph – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 127: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 127

IP reject

The IP reject sets lower level of the slope of the titration curve for ac-
ceptance of an inflection point. Use the IP reject value to eliminate
weak inflection points.
To set the appropriate IP reject value, display the derivative curve of
the titration and find the IP reject value as shown below:

Figure 22: IP reject

Enter in:

Parameters menu of a Monotonic IP or Dynamic IP method.

Range available:

1 to 999 (unit = mV/ml or pH/ml)

IP>1 (result
indicator)

Refer to "Result indicators", page 185

.

Iso pH

pH at which the electrode potential is no longer temperature depend-
ant. The Iso pH is an electrode characteristic supplied with every
Radiometer Analytical electrode.
Values are normally between 6.3 and 7.3 pH

Enter in:

Edit Electrode > Calibration parameters menu

Range available:

0.00 to 14.00 pH

IP reject 2

d Potential (or d pH)/dVolume

IP reject 1

Volume

IP reject 3

IP reject 4

IP1

IP2

IP3

IP reject 1: 3 Inflection points found (IP1, IP2 and IP3)
IP reject 2: 2 Inflection points found (IP2 and IP3)
IP reject 3: 1 Inflection point found (IP3)
IP reject 4: no inflection point found

Direct curve

Derivative curve