TitraLab 860 and 865 Reference Manual

Page 171

Reagent
calibration
stack

The reagent calibration stack defines individual data for each
standard solution beakers present in a reagent calibration sequence.

1.

In the Configuration menu, declare the model of sample changer
used (SAC80, SAC90, SAC850 or SAC950).

Refer to "Configuration menu", page 76

.

Depending on the model of sample changer used, enter the
sample changer configuration parameters.

Refer to "Sample changer", page 198

.

2.

In the Main window, select SAC Sequence or SAC method
for the working mode. This sequence or method must use
the reagent you want to calibrate. Edit the sequence or the
method if relevant,

see "Programming sequence", page 159

.

3.

Enter the Reagent window and press 1 Calibrate/Enter titre.

4.

Select the reagent to calibrate.

5.

Press 3 Calibration sequence.

Label the beakers indicating the running number in the
sequence, for example: 1/5, 2/5 etc.... and the name of the
standard solution.
Place the beakers in the numbered position on the sample
changer.
If rinses are programmed, position the corresponding rinse
beakers at the right places.

Refer to "Number of static rinses", page 146

.

Refer to "Number of dynamic rinses", page 144

.

You can print the calibration stack by pressing Print from the
calibration menu.

6.

Press Esc then run the sequence by pressing 1 Run calibration.

<1/5> means the first beaker over 5
programmed in the sequence. Use the LEFT/
RIGHT
arrows to review the other beakers in
the sequence.

The Standard name and sample amount are
entered in the Standard menu of the reagent
(ID and Standard amount parameters).

Enter the batch number of each standard
solution.