GLP-Archives
menu

The GLP-Archives (Good Laboratory Practice) command is available
in each window:

To access:

Sample results: enter Main window and press 5

Reagent calibration results: enter Reagent window and press 6

Electrode calibration results: enter Electrode window and press 6

Storage capacity:

Last 200 sample results.
Last 100 reagent calibration results.
Last 100 electrode calibration results.

When the GLP-Archives is full and a new result arrives, the
oldest result stored will be the first one removed.

Ground conflict

Ground conflict: External grounding defined in Setup/Configura-
tion and a metal electrode.

An external grounding is defined for the titration system cell in the
Configuration screen and a metal electrode is used by the method.

When a metal electrode is used, select TIM cell external
Gnd = No in the Configuration menu.

Help

Refer to "Check command", page 74

.

High (result
indicator)

Refer to "Result indicators", page 185

.

Press

✓ and enter the result number from where

you want to start visualising results, e.g. result
no. 9 over 46.

Press 1 to consult archives