Glp-archives menu, Ground conflict, Help – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 122: High (result indicator), See "glp-archives menu, Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 122
GLP-Archives
menu
The GLP-Archives (Good Laboratory Practice) command is available
in each window:
To access:
•
Sample results: enter Main window and press 5
•
Reagent calibration results: enter Reagent window and press 6
•
Electrode calibration results: enter Electrode window and press 6
Storage capacity:
Last 200 sample results.
Last 100 reagent calibration results.
Last 100 electrode calibration results.
When the GLP-Archives is full and a new result arrives, the
oldest result stored will be the first one removed.
Ground conflict
Ground conflict: External grounding defined in Setup/Configura-
tion and a metal electrode.
An external grounding is defined for the titration system cell in the
Configuration screen and a metal electrode is used by the method.
When a metal electrode is used, select TIM cell external
Gnd = No in the Configuration menu.
Help
Refer to "Check command", page 74
High (result
indicator)
Refer to "Result indicators", page 185
.
Press
✓ and enter the result number from where
you want to start visualising results, e.g. result
no. 9 over 46.
Press 1 to consult archives
Press 2 to consult the Global variables