Electrode system, Electrode type, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 106

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 106
Electrode
system
An electrode system comprises all the electrodes necessary to run a
method or a sequence of methods.
A method, consists of an indicating electrode, a reference electrode
and, if required a temperature sensor.
A sequence, can consist of several indicating electrodes.
When a method/sequence is run, the instrument prompts you
to connect or disconnect the electrodes that will be required to
run this method/sequence.
Electrode type
The electrode type is displayed with respect to the function selected
(
). The electrode type is defined when a new
electrode is created.
Refer to "Create electrode", page 82
The different electrode types are listed below:
Table 4: Electrode functions and types
If Combined pH is defined, the Titration Manager prompts
you to specify if it has a built-in temperature sensor.
If a Single electrode is defined, the Titration Manager
prompts you to define a reference electrode.
Type
Function
Single pH
pH
Combined pH (without temp. sensor)
pH
Single metal/redox
mV (i=0)
Combined metal/redox (without temp.
sensor)
mV (i=0)
ISE single
mV (i=0)
ISE combined (without temp. sensor)
mV (i=0)
Reference
Reference
Temperature sensor
T°C
Metal
Ground
Double metal
mV (i >0)