Parameter

RS 232 BAUD RATE

RS 232 BAUD Values: 300, 1200, 2400, 9600

Units

BAUD

RS 232 BAUD
RATE


RS 232 BAUD RATE

This parameter sets the host port communications baud rate. The higher the baud rate, the more

data bit transfer is increased which, generally speaking, results in faster communications. In an electrically
noisy environment, lower baud rates can be tried if communications fail at higher baud rates. In an
electrically noisy environment, the communications may not appear any faster using the higher baud rates
due to the increased number of message retries (corrupted messages are requested by the receiver to be re-
sent). Noise has a greater effect at higher baud rates because the duration of a noise pulse occupies a higher
ratio of time when compared with the duration of a signal pulse. Please ensure proper system grounding is
maintained as well as proper cable grounds and shields.
____________________________________________________________________
Parameter

RS 232 PROTOCOL

Protocol types: Sycon, ASCII

Units

None

RS 232
PROTOCOL TYPE


RS 232 PROTOCOL

This parameter sets the host port communications protocol type. The Sycon protocol has a higher

degree of integrity but is a proprietary protocol and will work with Sycon software only. The ASCII
protocol has lower integrity but is universally accepted. There is included with the STC-2000A a CD that
contains various software programs using both protocols.