STC-2000A DEPOSITION CONTROLLER

y

Used to stop the deposition process in non-sequencing mode. Non-sequencing mode does
not use a process program but rather a film, which is invoked by making it the next active
film and pressing the START key sequence to begin the deposition process. It has,
therefore, one implicit process. The Final Thickness Limit Trigger stops the selected film
deposition when the thickness reaches this user entered thickness value. This value is
analgous to the thickness value in a process program when in sequencing mode.
______________________________________________________________________

Parameter 8

Setpoint Time Lim

it

access:

SPTM

Value Range

0:00 - 99:59

Units

Min:Sec

The setpoint time limit parameter is used to provide a comparison point for the STC-
2000A timer setpoint event status.
______________________________________________________________________

Parameter 9

Soak 1 P

o

w

e

r Value

access:

PWR1 power condition power level 1

Value Range

0.0 - 100.0

Units

Percent

This parameter sets the deposition source power value to be used as the first level of soak
power in the pre-deposition cycle. This power level will be maintained for the duration of
the soak 1 timer. This level is typically used to out gas and condition the source material.
A value of 0.0% causes the rise/soak 1 phase to be skipped.
______________________________________________________________________


Parameter 10

P

o

w

e

r Ramp 1 Time

access:

RMP1 ramp time to power level 1

Value Range

0:00 - 99:59

Units

Min:Sec

This parameter sets the duration of the power ramp from zero power to the power level
set by the soak 1 power value parameter.
______________________________________________________________________

Parameter 11

P

o

w

e

r Soak 1 Time

access:

SKT1 soak time at power level 1

Value Range

0:00 - 99:59

Units

Min:Sec

This parameter sets the time that the source will remain at the soak 1 power level
following the completion of the ramp to this level.
______________________________________________________________________

Parameter 12

Soak 2 P

o

w

e

r Value

access:

PWR2 power condition power level 2

Value Range

0.0 - 100.0

Units

Percent

This parameter sets the deposition source power value to be used as the second level of
soak power in the pre-deposition cycle. This power level will be maintained for the
duration of the soak 2 timer. A value of 0.0% skips the rise/soak 2 phases.
______________________________________________________________________



Parameter 13

P

o

w

e

r Ramp 2 Time

access:

RMP2 ramp time to power level 2

Value Range

0:00 - 99:59

Units

Min:Sec

SECTION 6.XX