STC-2000A DEPOSITION CONTROLLER

y

Value Range

5000000.0…6000000.0, 0.0 if bad crystal

Units

Hz

This is the measured frequency of the active sensing crystal.
______________________________________________________________________

The following parameters (14-61) are the values as reported by physically present and enabled source
sensor card[s].
______________________________________________________________________

Parameter 14

Crystal frequency from Sensor 1

access index:

FREQ1 Crystal frequency from Sensor 1 in centiHz (0.01Hz) resolution.

Value Range

500,000,000-600,000,000

Units centiHz

(0.01Hz)

Crystal frequency from Sensor 1 in centiHz (0.01Hz) resolution [6 MHz looks like 600,000,000].

Associated card needs to be physically installed and enabled before a reboot.

______________________________________________________________________

Parameter 15

Crystal frequency from Sensor 2

access index:

FREQ2 Crystal frequency from Sensor 2 in centiHz (0.01Hz) resolution.

Value Range

500,000,000-600,000,000

Units centiHz

(0.01Hz)

Crystal frequency from Sensor 2 in centiHz (0.01Hz) resolution [6 MHz looks like 600,000,000].

Associated card needs to be physically installed and enabled before a reboot.

______________________________________________________________________

Parameter 16

N.A.

______________________________________________________________________

Parameter 17

N.A.

______________________________________________________________________

Parameter 18

N.A.

______________________________________________________________________

Parameter 19

N.A.

______________________________________________________________________

Parameter 20

N.A.

______________________________________________________________________

Parameter 21

N.A.

______________________________________________________________________

Parameter 22

source 1 power level

access index:

POWR1 power level, in units of .1 percent, 0.. 1000, of source 1.

Value Range

0 to 1000

Units

.1

Percent

source 1 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________

Parameter 23

source 2 power level

access index:

POWR2 power level, in units of .1 percent, 0.. 1000, of source 2.

Value Range

0 to 1000

Units

.1

Percent

SECTION 6.XX