INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual

Page 228

Advertising
background image

p

STC-2000A DEPOSITION CONTROLLER

y

This parameter is used to indicate that the STC-2000A is in a stopped state. The following returned values
are:

0

Stopped From A Power Up Condition

1

Stopped From A Crystal Bad Condition

2

Stopped From The I/O System Or Computer

3

Stopped From The Front Panel Key

4

Stopped From A Max. Power Condition

5

Stopped From The Hand Controller Button

6

Not

Stopped

This is a read only parameter.
access: SMODE Stop mode, set when a stop event declared, set to one of the following STPCD_ types.

0 STPCD_CLEAN stopped at power up, no reason yet.

1 STPCD_GONE when stop condition cleared.
2 STPCD_XBAD crystal head or heads bad.

3 STPCD_RMOTE remote from I/O system.
4 STPCD_FPNL front panel.
5 STPCD_MAXP max power condition persisted too long.

6 STPCD_PEND remote pendant stop.

7 STPCD_UIFLST user interface presence lost.
8 STPCD_MEASCRIT

critical measurement failure, comm lost to S4 card required for usage per config.

9 STPCD_IOCRIT critical I/O failure, comm lost to I/O or a detectable/unrecoverable failure.

10 STPCD_BHIE_ERR Bus Hold Integrity error failure caused.

11 STPCD_INDEXER_FAIL Rotator failed to signal done in time.

12 STPCD_INVALID_SRC Invalid source programmed in recipe.

13 STPCD_INVALID_SNS Invalid sensor channel in recipe.

14 STPCD_IO_AX1 Auxiliary I/O stop 1.

15 STPCD_IO_AX2 Auxiliary I/O stop 2.

16 STPCD_IO_AX3 Auxiliary I/O stop 3.

17 STPCD_IO_AX4 Auxiliary I/O stop 4.

18 STPCD_QUALDLY Shutter delay quality failure

19 STPCD_QUALDEP Deposition quality failure.

______________________________________________________________________

Parameter 6

Active Film Number

Value Range

1 To 50

Units Film

Number

This parameter returns the current active film number. This is a read only parameter.
access: AFPN active film processing # 1..50.
______________________________________________________________________

Parameter 7

Running A Process (Process Running Flag)

Value Range

0,1

Units

False / True

This parameter indicates that a process is running. A returned value of 1 indicates that a process is running.
This is a read only parameter.
access: RNMD true when sequencer running a process, false(0)/true(1) false when at rest.

0 RNMD_FALSE

1 RNMD_TRUE

______________________________________________________________________

Parameter 8

Substrate Shutter Status

Value Range

0,1

Units Closed,

Open

This parameter indicates if the substrate shutter is open or closed. A returned value of 0 indicates that the
shutter is closed and a value of 1 indicates that it is open. This is a read only parameter.

SECTION 6.XX

page 228 of 292