STC-2000A DEPOSITION CONTROLLER

y

Read or set the process number that is currently active. The set command is inhibited if a process is
running.
______________________________________________________________________

Parameter 3

Process Run Number

access value:

RUNNO run number

Value Range

0 To 9999

Units

None

Read or set the run number. The set command is inhibited if a film is running.
______________________________________________________________________

Parameter 4

Password Lock # - Parameters And Executive

access value:

LOCKN param & exec lock number

Value Range

0 To 9999

Units

None

This allows certain areas of the STC-2000A to be restricted to those people knowing the password code. If
this is programmed with a non-zero value the value enables the password function and is itself the
password. Whenever accessing a menu that can change a parameter, the STC-2000A will ask for the
password.

______________________________________________________________________

Parameter 5

Password Lock # - I/O And Communications

access value:

COMLOK I/O and communications lock code

Value Range

0 To 9999

Units

None

This allows certain areas of the STC-2000A to be restricted to those people knowing the password code. If
this is programmed with a non-zero value the value enables the password function and is itself the
password. Whenever accessing a menu that can change an I/O programming parameter or a
communications parameter, the STC-2000A will ask for the password.
______________________________________________________________________

Parameter 6

NA

access value: OBSO Was control loop polarity

______________________________________________________________________

Parameter 7

Control F.S. Volt

Control Loop Full Scale Voltage

access value:

OBSO Was control loop full scale volts choice

Value Range

0 , 1 or 2

Units

0 = 2 Volts, 1 = 5 Volts, 2 = 10 Volts

This scales the control voltage of the STC-2000A to match your power supply. While most supplies are set
to -10 volts for full power, not all are. This allows 100% power from the STC-2000A to be either 2, 5, or
10 volts.
______________________________________________________________________

Parameter 8

Rate Direction

access value:

OBSO Was rate direction, positive, negative

Value Range

0,1

Units

0 = Positive, 1 = Negative

This parameter sets the STC-2000A for either a deposit mode (positive) or an etch mode (negative).
______________________________________________________________________

Parameter 9

Recorder Function

access value:

ARFNC analog recorder function select, ARFC_class RATE/RDEV/PWRL THIK/REMOTE/IOACC/NONE

Value Range

0 To 6

Units

0=

Rate,

1=

Rate Deviation,

2=

Power,

3=

Thickness,

4=

Computer Remote, 5=I/O Control, 6=Off

SECTION 6.XX

page 216 of 292