STC-2000A DEPOSITION CONTROLLER

y

Parameter 27

Max. Power Dwell

access:

MPTM max power time interval

Value Range

0:0 - 99:59

Units Min:Sec
This parameter is associated with the max. power abort switch parameter and sets the
allowable duration of a maximum power condition before an abort will occur.

______________________________________________________________________

Parameter 28

Shutter Delay Mode

Enable

access:

SDMD shutter delay mode, on/off

Value Range

0,1

Units

0 is Off, 1 is On

This switch type parameter will allow the shutter delay parameters to become active and
the mode functional on a individual film basis. If set to Off, the parameter window cursor
will not access the associated parameters. If set to On, the delayed shutter operation will
be performed as specified by the following parameters.

______________________________________________________________________

Parameter 29

Shutter Delay Time-Out

access:

SDTM time to achieve control without alarm

Value Range

0:01 - 99:59

Units

Min:Sec

This parameter sets a time limit for the delayed shutter operation to be completed. The
time period begins at the beginning of the shutter delay cycle. The sensor is exposed to
evaporant with the substrates remaining shielded. When rate control is established to the
programmed accuracy, three things happen. The substrate shutter is opened, accumulated
thickness set to zero, and the programmed deposition profile is then started. If the
specified rate control accuracy cannot be met within this time limit, the deposition will be
aborted.

______________________________________________________________________

Parameter 30

Shutter Delay Quality

access:

SDAC shutter delay accuracy factor

Value Range

1 - 50

Units

Percent

This parameter sets the rate control accuracy that must be established for a period of five
seconds in order to complete the delayed shutter sequence. The maximum accuracy set
able is 1 percent of the desired setpoint. However, this is internally limited to 1 Å/s.

______________________________________________________________________

Parameter 31

Rate Sampling

access:

XSMD crystal sample mode, off/timed/intel

Value Range

0 To 2

Units

0 is Off, 1 is Timed Mode, 2 is Intelligent Mode

This switch type parameter will allow the rate sampling parameters to become active and
the mode functional on an individual film basis. If set to Off the parameter window
cursor will not access the associated parameters. If timed mode is selected, the crystal
sensor will be exposed to evaporant for the time specified by the sample dwell parameter,
and will sample at the interval specified by the sample interval parameter. If intelligent
mode is selected, then the duration of the sample will only be long enough to establish
the desired accuracy.

______________________________________________________________________

Parameter 32

Sample Interval

SECTION 6.XX