STC-2000A DEPOSITION CONTROLLER

y


Options/Info, page 5 of 6

Options/Info, page 6 of 6

Displays live frequency measurements from

Displays Memory Module information if the

the installed/enabled source sensor cards.

optional memory module interface bd. and the

If no crystal/oscillator w/installed/enabled, freq. = zero

memory module are installed and the memory

[only screen w/ live update]

module is enabled. See section 3.20 for details.



This information can be useful to the factory if there is a problem. Information also appears after product
power-on during initialization. The initialization process (for diagnostic purposes) is as follows (with
typical LCD screen information):

1.

LEDs:

user programmable LEDs non-illuminated. Power LED illuminated.

LCD

display:

screen:

clear

backlight:

on

LEDs:

user programmable LEDs on (test)

2.

LCD display:

[one beep with the appearance of each text line]






Beeper: beeps with the appearance of each line of above text (5 beeps)

3.

LCD display:

screen clear

LEDs:

user

programmable

LEDs

non-illuminated

Beeper:

2

Lo/Hi

beeps

4. LCD

display:








5.

LCD display:

typical RunTime Screen

CST

APP LOADER V01.00(BUILD 749C)
----------------------------------------------
>APP FOUND:SECTOR E BUILD F904
>VERIFY APP IN PROGRESS
..............................................
>APP OK
>LAUNCHING APPLICATION...

CSUB: VERSION TKAPA03.15 BUILD F904
SYNCING WITH CSUB...60

LEDs:

controlled by I/O programming while running


check sum validation
Check sums are used to validate the integrity of the user programmable functions for the STC-2000A. Of
the programs that use check sums, there are three categorical types: one for the film parameters (Main
Menu: Review Film key), another for the process steps (Main Menu: Review Processes key), and finally