STC-2000A DEPOSITION CONTROLLER

y

This parameter assumes there is a shuttered head connected and that Rate Sampling can be used if

the film parameters are programmed for such. (Without the shutter, deposit would be continuous and negate
the advantage of this feature.) This feature extends the apparent life of the sensing crystal. It is used for
thick coatings or in systems where it is not possible to break vacuum to change the sensing crystal.
Depending on various factors in your particular system, the apparent crystal life can be extended up to 50
times by using the rate sampling feature of the STC-2000A. Five film parameters which pertain to this
feature are described in (Section 2.13).
____________________________________________________________________
Parameter

SAMPLE INTERVAL

Process Time; Range is 0:01 - 99:59

Units

Min:Sec


Sample Interval Time

For either mode of operation this parameter will select the sampling interval of the sensor.

____________________________________________________________________
Parameter

SAMPLE DWELL TIME

Process Time; Range is 0:01 - 99:59

Units

Min:Sec


Sample Dwell Time

For the Timed Mode selection this parameter sets the duration of sensor exposure.

____________________________________________________________________
Parameter

SAMPLE QUALITY

Quality Value; Range is 1 - 50

Units

Percent


Sample Quality

In the Intelligent Mode this parameter sets the control quality required to terminate the exposure

period of the sensor.
____________________________________________________________________
Parameter

SAMPLE ALARM TIME

Process Time; Range is 0:0 - 99:59

Units

Min:Sec


Sample Alarm Time

In the Intelligent Mode this parameter sets the maximum allowable time allowed to complete the

sensor exposure and control operation. The STC-2000A will treat this as a crystal failure when this limit is
exceeded.
____________________________________________________________________
Parameter

FILM FAIL MODE

Meas. Processing: select 1 of 2 choices;

Abort

If

Fail

Complete On Time Power

Units:

None

Enabling Measurement Fail Processing

This parameter allows the user to select the system execution path if it is no longer possible to get

reliable measurement information from the sensor system. This is the last resort choice that will be made
after all other system configuration options have been exhausted. Choice 1) Abort If Fail, will abort the
deposition and set the output power to zero. Choice 2) Complete On Time Power, will set the unit into a
time power completion mode. In this mode the last reliable control power level will be maintained and
thickness accumulation will occur in a simulated fashion until the final thickness setpoint is reached. Post
deposition processing then will occur as programmed.
____________________________________________________________________
Parameter

CTL LOOP QUAL L

Select Value; Range is 0 - 9

Units

None

SECTION 3.XX

page 81 of 292