Films and processes, Film edit/review mode, Stc-2000a deposition controller – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual

Page 69: Runtime information

Advertising
background image

p

STC-2000A DEPOSITION CONTROLLER

y

SECTION 3.4

Films and Processes

STC – 2000A Film Process Data Storage

FILM # 1 PARAMETERS

FILM # 2 PARAMETERS

FILM # 3 PARAMETERS

FILM # 4 PARAMETERS

FILM # 5 PARAMETERS

FILM # 6 PARAMETERS

FILM # 7 PARAMETERS

FILM # 8 PARAMETERS

FILM # 50 PARAMETERS

PROCESS LIST # 1

PROCESS LIST # 2

PROCESS LIST # 3

PROCESS LIST # 4

PROCESS LIST # 5

PROCESS LIST # 6

PROCESS LIST # 7

PROCESS LIST # 8

PROCESS LIST # 9

RUNTIME

INFORMATION

ACTIVE PROCESS 1-9

USE FILM #

LAYER #

PROCESS RECIPE LIBRARY

FILM RECIPE LIBRARY

2 AUTO 2 0.400

1 AUTO 3 0.030

3 WAIT 3 0.600
4 AUTO 2 0.250

Step Mode Film# Thickness KA

UP
TO

99

END

1

3

1

Figure- 3.8: Film & Process Storage.


Films & Processes Description

A deposition process is controlled by the parameters programmed by the user in the film

parameter and process sequencing lists. Figure 12.8, illustrates the film parameter and process data storage
and the relationship to the RunTime mode. The figure shows a process (Process 1) sequencing through the
layers (currently on layer 1) using the parameters of Film 3. The ending thickness for this layer is
determined by the final thickness for step 1 (layer 1) programmed in the process list. Each of the nine
processes can have a different set or arrangement of any or all of the 50 films. A process list can
repetitively use a film program for different layers (steps) as in the case for layer 1 & 3 of Process 1. The
ending thickness for each layer is independent of the film parameters and determined by the value
programmed in the process list.


SECTION 3.5

Film Edit/Review Mode

Editing Films

The STC-2000A has storage for 50 sets of film parameters (50 films). Depending on the system

and hardware configurations, up to 42 parameters per film can be stored. On the MAIN MENU, the Review
Film
key provides entry into the menu area containing all film edit/review menus. An unprogrammed unit
will start with film 1 selected. When the REVIEW FILM key pressed, this brings into view the film
parameter list, which was last selected as the film to review or edit using the numeric entry screen. The film
number that is currently being viewed and can potentially be edited, is indicated on the top line of the
display (menu path/status/ID line) in the form EDIT x, where x is the film now open for viewing/editing.
If the film number that is displayed when is the REVIEW FILM key pressed is not the film needed for
viewing/editing, press the GO2 FLM (go to film) key to bring forth the numeric entry screen. This is simply
a vehicle used to go to the film number as indicated by the entered number. This may or may not be the
active film, that is, the current film that is in use. The Active Film number value can be seen, as was the
edit film number, on the top line of the REVIEW FILM menu screen in the form ACTIVE x, where x is the
film now active for a starting process. The Active Film number value is also indicated on the RunTIme
screen [ FILM __ ]. To edit or review a different film while on the REVIEW FILMS menu , press the GO2
FLM
key again and change digit[s] as required. The alternate film number will be selected whereupon
editing/reviewing can commence on the subsequent menu screens as well as on the REVIEW FILM menu.

SECTION 3.XX

page 69 of 292

Advertising