STC-2000A DEPOSITION CONTROLLER

y

255 ALL sources overiden to zero power.
______________________________________________________________________

Parameter 22

At Rate (

Deposition is at Desired Rate)

Value Range

0, 1

Units False,

True

This parameter is TRUE if the measured rate is equal to the desired rate, else (0) not at desired rate.
access: ATRATE Deposition was at rate, NOT (0), YES (1). Set by measurement once rate is > desired.
Only cleared in I/O program if used by customer. Better to use event code for ATRATE.
______________________________________________________________________

Parameter 23

Allow Zeroing

Value Range

0, 1

Units False,

True

Check if Zeroing allowed when entering deposit. If this parameter is true (1) then Zeroing is inhibited on
entering a deposit.
access: ZERO Allow normal zeroing (0=ZERO_NORM) or inhibit it (1=ZERO_SKIP) when entering
deposit.
______________________________________________________________________

Parameter 24

Crystal Test Mode

Value Range

0, 1

Units False,

True

If this parameter is true, a dual sensor configured STC-2000A will successfully start a film with only one
sensor operational (the limp along mode).
access: SCUP_SEES_UIFC Set true when UIFC detected.
______________________________________________________________________

Parameter 25
access: MEAS_DEAD Set true when measurement system fatal failure detected. This is a hardware
problem which is unrecoverable without recycling power and possibly repairing unit.
______________________________________________________________________

Parameter 26
access: UIF_DOGDIE Set true when UIFC dog times out and stop occurs.
______________________________________________________________________

Parameter 27
access: TEST_MODE Set true when the test/simulate mode is active.
______________________________________________________________________
Parameter 28
access: RTRSYNC Rotator Synchronization needed, 1 per source.
______________________________________________________________________
Parameter 29
access: ACTSRC Active source being controlled, 1..8.
______________________________________________________________________
Parameter 30
access: NON_SEQ Non-sequencing when true.
______________________________________________________________________
Parameter 31
access: SA1_POCKET source Channel 1 pocket, 0-63
______________________________________________________________________
Parameter 32
access: SA2_POCKET source Channel 2 pocket, 0-63
______________________________________________________________________
Parameter 33

SECTION 6.XX