Remote inputs for factory installed i/o program, Rs-232 serial communications interface – INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual

Page 154

Advertising
background image

p

STC-2000A DEPOSITION CONTROLLER

y

Remote Inputs for Factory Installed I/O Program

The standard STC-2000A is equipped with eight remote input functions. These inputs are
intended to be activated in a number of ways depending upon the previously described jumper
placement (see figure 3.3). The signal can be continuously applied or be pulsed. The STC-
2000A can be programmed to respond to the leading or trailing edge of this pulse. The

minimum pulse width required is 15 ms.

All eight of the inputs can be programmed by the user to suit the

system requirements. It is assumed that the INPUT card PCB is in slot 1 (see figure 3.1 for I/O slot
designations). Initially, the first four inputs are pre-programmed at the factory to the following four
functions:
Factory Remote Inputs
Input 1

START - Functionally identical to Front Panel START key with the STC-2000A in the
ready state. Activating this input with the STC-2000A in either the ready or stopped state
will begin the process cycle.

Input 2

STOP -Functionally identical to Front Panel STOP key. Activating this input will
terminate the deposition process.

Input 3

Final Thickness Trigger - Activation of this input causes the unit to respond as if the final
thickness has been reached.

Input 4

Zero Film Thickness.

External P.S. Activation 3.5 to 40 VDC , AC voltage from 5 to 30 Volts rms

Minimum

Current 2

mA

SECTION 4.10

Specifications

SECTION 4.11

RS-232 Serial Communications Interface

This connection provides a serial data communications link to the STC-2000A. Four standard baud rates
can be selected: 300, 1200, 2400, and 9600. They are selected via the configuration menu. The serial
interface communications protocol is also selectable. The choices are either the Sycon format or ASCII
protocol. The electrical specifications correspond to the EIA RS-232 standard.

RS-232 Connections

STC-2000A Type: 9 Pin "D" male

Mating Connector: AMP 205203-1 Or Equivalent (Sycon#s: 404-028, 404-019)

STC-2000A

Function

Pin Number

RS-232 Received Data

2

RS-232 Transmitted Data

3

RS-232 Signal Ground

5

Cable Shield

Shield Case


SECTION 4.XX

page 154 of 292

Advertising