STC-2000A DEPOSITION CONTROLLER

y

Units

map#

This parameter selects the map that is to be associated with film that invokes it.

______________________________________________________________________

Parameter 45

POCKET SELECT

access:

PSEL Pocket select 0..63

Value Range

0 – 63

Units

Pockets

This parameter selects the type of information to be displayed on the RunTime graphics
screen.

______________________________________________________________________


Parameter 46

ETCHING MODE

access:

FRATDR

Film Rate Direction/EtchMode if true, otherwise deposit.

Value Range

OFF/ON

Units

None

When ON, this parameter selects the deposition controller perspective such that a
negative (etching) process will be indicated with positive values and that setpoints, etc.
will trigger based on these values. When OFF deposition will be indicated with positive
values.

______________________________________________________________________

Read, Write Film Parameters Examples

Command Examples

Command

A1 25 or A 1,25 or A 1 25

Response A0.0
Description

To Request Film 1, Parameter 25 The Max. Power Limit Parameter. The Current
Parameter Setting, In This Case 0.0 %.


Command
A9

18

Response A0
Description

To Request Film 9, Parameter 18, Rate Ramp Mode. Rate Ramp Mode is Off.


Command

B1 25=50.2 or B 1,25 = 50.2

Response A
Description

To Set Film 1, Parameter 25 (The Max. Power Limit Parameter) To 50.2 %. Message
Understood And Accepted.


Command B1

25=100.4

Response H
Description

To Set The Parameter To A Value Exceeding 100%, Which is Illegal! Response is
Signifying An Illegal Data Value Being Used, command ignored.


Table 6.4: Group 2 Examples.
______________________________________________________________________




______________________________________________________________________

Read, Write System Parameters

SECTION 6.XX

page 214 of 292