INFICON STC-2000A Thin Film Deposition Controller Operating Manual User Manual

Page 77

Advertising
background image

p

STC-2000A DEPOSITION CONTROLLER

y

This parameter sets the deposition source power value to be used as the first level of soak power in

the pre-deposition cycle. This power level will be maintained for the duration of the soak 1 timer. This
level is typically used to out gas and condition the source material. A value of 0.0% causes the
RISE/SOAK 1 phases to be skipped.
____________________________________________________________________
Parameter

PWR RAMP 1 TIME

Elapsed Time; Range is 0:00 - 99:59

Units

Min:Sec

Ramp Time To Soak 1

This parameter sets the duration of the power ramp from zero power to the power level set by the

soak 1 power value parameter. The source is typically heated slowly enough to permit outgassing and melt
conditioning without causing actual deposition to take place.
_____________________________________________________________________
Parameter

PWR SOAK 1 TIME

Elapsed Time; Range is 0:00 - 99:59

Units

Min:Sec

Soak 1 Time

This parameter sets the time that the source will remain at the soak 1 power level following the

completion of the ramp to this level.
____________________________________________________________________
Soak 2 Power Value
Parameter

SOAK 2 PWR VALUE

Percentage Value; Range is 0.0 - 100.0

Units

Percent


This parameter sets the deposition source power value to be used as the second level of soak

power in the pre-deposition cycle. This power level will be maintained for the duration of the soak 2 timer.
This level is typically used to further out gas and condition the source material and actually begin
depositing material at a low rate. A value of 0.0% causes the RISE/SOAK 2 Phases to be skipped.
____________________________________________________________________
Parameter

PWR RAMP 2 TIME

Elapsed Time; Range is 0:00 - 99:59

Units

Min:Sec

Ramp Time to Soak 2

This parameter sets the duration of the power ramp from the soak 1 power value or the idle power

value to the value set by the soak 2 power parameter. The source is typically heated slowly enough to
permit further outgassing and melt conditioning and actual deposition may begin during this time period.
____________________________________________________________________
Parameter

PWR SOAK 2 TIME

Elapsed Time; Range is 0:00 - 99:59

Units

Min:Sec

Soak 2 Time

This parameter sets the time that the source will remain at the soak 2 power level following the

completion of the ramp to this level. At the end of this time the deposition phase begins. An external input
can be programmed using the programmable I/O to hold the film at soak until released (Soak Hold).
____________________________________________________________________
Parameter

SOAK 3 PWR VALUE (IDLE PWR)

Percentage Value; Range is 0.0 - 100.0

Units

Percent

Idle Power Value

This parameter determines the final (idle) power setting at the completion of the entire deposition

profile. If this power level is zero, a START command will begin a new power profile at rise 1, an idle
power setting other than zero will cause the power profile to begin at rise 2 after a START.
_____________________________________________________________________

SECTION 3.XX

page 77 of 292