16 stdadd auto, 1 stdadd auto - overview, Stdadd auto – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1001

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

985

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

0.1 ... 999999.0 s

Default value

2.0 s

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously.

Range

-20.0 ... 150°C

Default value

25.0

Switch off stirrer during measurement

Switch off stirrer during measurement

on | off (Default value: off)

If this option is activated, then the stirrer is switched off during the meas-
urement. In this case the following two parameters can be edited:

Stir before measurement

Stirring time between end of dosing and measurement.

Range

0 ... 999999 s

Default value

0 s

Pause before measurement

Waiting time before the measurement during which no stirring takes place.

Range

0 ... 999999 s

Default value

0 s

5.6.4.16

STDADD auto

5.6.4.16.1

STDADD auto - Overview

Dialog window: Method

STDADD auto Properties... STDADD auto - 'Command

name'

Command for Standard addition with automatic addition of the standard
addition solution from a dosing device in such a way that a constant poten-
tial difference results.

Devices
This command can be executed with the following devices:

Advertising