4 set upol - control parameters, Control parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 731

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

715

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.4.6.4

SET Upol - Control parameters

Tab: Method

SET Upol Properties... Control parameters

Command name

Name of the command.

Input

25 characters

The control parameters can be set for each endpoint separately.

Endpoint 1

EP1 with

Measured value for the first endpoint.

Range

-200.0 ... 200.0 µA

Selection

off

Default value

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical, fast titrations.