5 cal cond - data conductivity standard, 9 cal spec, 1 cal spec - overview – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1047: Cal spec

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1031

Name standard

List box with conductivity standards. The data of one of these standards is
used to calculate the temperature coefficient for the specified reference
temperature.

Selection

'Conductivity standard'

5.6.5.8.5

CAL Cond - Data Conductivity standard

Dialog window: Method

CAL Cond Data Conductivity standard

Current temperature

Display of the currently measured temperature of the conductivity standard.

Data transmission from standard
Input of the data of the conductivity standard, which should be as close as
possible to the Current temperature. This data is as a rule contained in
the Data Sheet/Certificate for the conductivity standard.

Reference temperature

Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.

Range

-20.0 ... 150.0°C

Default value

20.0

Conductivity

Input of the conductivity of the standard at the reference temperature.

Range

0.00000 ... 2000.00 mS/cm

Default value

1.0 mS/cm

Temperature coefficient

Temperature coefficient of the standard with the reference temperature.

Range

0.00 ... 9.99 %/°C

Default value

1.00 %/°C

5.6.5.9

CAL Spec

5.6.5.9.1

CAL Spec - Overview

Dialog window: Method

CAL Spec Properties... CAL Spec - 'Command name'

Command for wavelength calibration.

Devices
This command can be executed with the following devices:

Spectrometer: Avantes

Advertising