7 Devices

tiamo 2.1 (ProcessLab)

■■■■■■■■

1427

Action
The actions can only be edited if monitoring is switched on.

If during monitoring it is found that the validity period has expired then one
of the following actions will be triggered automatically at Start test.

Selection

Record message | Display message | Stop deter-
mination

Default value

Display message

Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can choose whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the deter-
mination.
Stop determination
The running determination will be automatically terminated. The follow-
ing message must be confirmed with [OK].

7.16.5

Establish a connection to the RS-232 device

Dialog window: Configuration

Devices Properties RS-232 Connect

In order to establish a connection to a RS-232 device the button [Con-
nect]
has to be pressed on the tab RS-232.

If then a message is entered in the field besides the button [Send] and is
sent, the message received by the device appears in the large text field.

Text field

The messages sent to and received by the device are shown in the large text
field.

Input field

Input field for a message to be sent to the device.

[Send]

Sends the message shown in the input field to the device.