5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

541

Identification

Description

.MI.TIM

Time for the minimum measured value in s

.MI.VOL

Volume for minimum measured value in mL

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)

.NMP

Number of measuring points in the measuring point list

.SLO

Electrode slope of the sensor used for the command (in
mV for ISE sensors)

.SME

Start measured value (measured value after processing
the start conditions) in the unit of the measured value

.STE

Start temperature (temperature after processing of the
start conditions) in °C

.STY

Type of stop with which the command was stopped:
1 = normal;0 = manual or after error

.SVA

Start volume absolute (volume that was added accord-
ing to the start condition "start volume") in mL

.SVM

Start volume measured value (volume which was added
according to the start measured value given in the start
conditions) in mL

.SVS

Start volume slope (volume that was added according
to the start condition "start slope") in mL

.SVT

Total start volume (volume that was added in depend-
ence on all three start conditions) in mL

.TITER

Titer value of the solution used for the command

5.6.3.3.2.2

MET pH - General/Hardware

Tab: Method

MET pH Properties... General/Hardware

Command name

Name of the command.

Input

25 characters

The general parameters for the control device, the dosing device, the sensor
and the stirrer are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.