4 meas ipol - evaluations, Evaluations – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 910

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

894

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrino
Range

-170.0 ... 500 °C

Default value

25.0 °C

5.6.4.4.4

MEAS Ipol - Evaluations

Tab: Method

MEAS Ipol Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves can
be activated and defined on this tab.

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other measured
quantities will be interpolated with a fix measured quantity (Measured
value
or Time) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.4.4.6, page 898).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.4.6, page 898).

[Delete]

Deletes the selected line.