5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

779

Range

1 ... 999 µg/min

Default value

5 µg/min

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Range

0 ... 999 µg/min

Default value

5 µg/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is only
shown for Stop criterion = time.

Range

0 ... 999 s

Default value

10 s

5.6.3.6.5

KFC - Titration parameters

Tab: Method

KFC Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.

Extraction time

The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). The entry of an extraction time may be
advisable, e.g. for the titration of sparingly soluble samples.

Range

0 ... 999999 s

Default value

0 s

Temperature

Manually entered titration temperature.

756, 831
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

851, 852
Range

-20.0 ... 150.0 °C

Default value

25.0 °C