5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1072

■■■■■■■■

tiamo 2.1 (ProcessLab)

device, it will automatically be reduced during dosing to the largest possible
value.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

maximum mL/min

Note

The filling rate should be reduced for viscous liquids.

5.6.6.3.5.5

DOS U - Monitoring

Tab: Methods

DOS U Properties... Monitoring

Command name

Name of the command.

Input

25 characters

The following monitoring processes can be switched on and defined on this
tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is switched on then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

-2000 mV