7.19 IO controller

■■■■■■■■■■■■■■■■■■■■■■

1446

■■■■■■■■

tiamo 2.1 (ProcessLab)

>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in 10
ms
Example:
Parameter 1
= 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on 225
ms, switched off 75 ms

Range

-214783647 ... 214783647

Default value

0

7.19.6

Properties - Analog inputs

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Analog inputs

On the registration card Analog inputs all inputs detected are automatically
listed in a table. For each analog input the settings can be modified.

Port

Display of the internal description of the port.

Port description

Display of the port description.

Type

Display of the port type.

Low level (0%)

Display of the lowest analog input value with unit.

High level (100%)

Display of the highest analog input value with unit.

0% Mapping

Measured value to be used when the Low level (0%) is adjacent as input
value, with unit.

100% Mapping

Measured value to be used when the High level (100%) is adjacent as
input value, with unit.

[Edit]

Opens the dialog Properties Analog input in order to edit the selected
port.

[Port description]

Opens a menu in order to import or export port descriptions.

Advertising