2 options - general, 3 options - save – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1247

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.1 (ProcessLab)

■■■■■■■■

1231

6.3.4.2

Options - General

Tab: Configuration

Tools Options Options General

Dialog language

Dialog language

Selection of the dialog language.

Selection

German | English | additional languages
(dependent on installed language patches)

Default value

English

Note

For the altered setting to become effective the program must be restarted.

Emergency stop button
If this option is enabled then the emergency stop button will be shown in
all program parts.

Selection

on | off

Default value

off

This button can be moved to any position with the left-hand mouse button
pressed down and is always shown right up at the top on the screen, both
in the program window and outside it on the Windows desktop. A mouse
click on this button immediately stops all running determinations on all acti-
vated workplaces. When this happens, all devices are stopped (including
pumps; any device commands that may have been started will be carried
out to their ends, e.g. PREP with Dosino) and the exit track will be started.
The emergency stop is effective even if no user is logged in, e.g. when the
user has been logged out automatically.

The emergency stop has no effect on running actions in the manual control,
each of which must therefore be stopped with <Stop>.