4 kft upol - control parameters, Control parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 773

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

757

Titrino
Range

0.00 ... 999.99 mL

Default value

0.00 mL

Dosing rate

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.

Titrando
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.5.3.4

KFT Upol - Control parameters

Tab: Method

KFT Upol Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Endpoint

Ep with

Measured value for the endpoint.

Range

-200.0 ... 200.0 µA

Default value

25.0 µA

Selection

off