5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1056

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrando, 855
Range

0.1 ... 999999 s

Default value

5.0 s

Titrino
Range

1 ... 999999 s

Default value

5 s

5.6.6.3.4.4

DOS pH - Stop conditions

Tab: Method

DOS pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the dosing are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the dosing.

Stop volume

Stops when the given volume has been added after the start of the dosing.
The stop volume should be adapted to suit the sample weight or the titration
vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100.000 mL

Selection

off

off
Means that no stop will take place.

Titrino
Range

0 ... 9999.99 mL

Default value

100.00 mL

Selection

off

off
Means that no stop will take place.

Stop time

Stops when the preset time has elapsed since the start of the dosing.

Range

0 ... 999999 s

Selection

off

Default value

off