5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

742

■■■■■■■■

tiamo 2.1 (ProcessLab)

5.6.3.5.2.6

KFT Ipol - Stop conditions

Tab: Method

KFT Ipol Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.00 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Stops when the preset time (including start conditions) has elapsed since the
start of the titration.

Titrando
Range

0 ... 999999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is to be filled after titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used.

Titrando
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum