8 kft ipol - additional evaluations, Additional evaluations – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 761

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

745

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.

5.6.3.5.2.8

KFT Ipol - Additional evaluations

Tab: Method

KFT Ipol Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for the evaluation of the titration curves
can be activated and defined on these tabs:

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value
, Volume or Time) for the fix endpoint from the measuring
point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.5.2.10, page 748).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.5.2.10, page 748).