5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1143

Selection

min | s

Default value

s

5.6.9.6.3

TRANSFER - Transfer command

Dialog window: Method

TRANSFER Properties... Transfer command

Name

Freely definable name for the transfer command.

Input

50 characters

Action

Selection

Read | Write | Wait for | Poll

Default value

Read

Read
The character string defined in Command (optional) will be sent to the
device and the answer of the device is expected within the Timeout
defined for the device settings and evaluated.
Write
The character string defined in Command (optional) will be sent to the
device.
Wait for
The character string defined in Command (optional) will be sent to the
device, after which there is a waiting period until a received character
string corresponds to the regular expression in Comparison. If this
answer is not received within the Timeout defined in the command,
then the command is canceled.
Poll
The character string defined in Command (optional) will be sent to the
device; an answer is expected from the device and evaluated within the
Timeout defined for the device settings. This action will be repeated
until the answer of the device is identical to the regular expression in
Comparison. If this answer is not received within the Max. waiting
time
defined in the command, then the command is canceled.

Command

Character string to be sent to the device. Details can be obtained from the
documentation for the respective device.