3 method window - live modifications, Method window - live modifications, Modify parameters live – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 181

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Workplace

tiamo 2.1 (ProcessLab)

■■■■■■■■

165

Selection

Fit in window

Default value

Fit in window
Adjust to both height and width of the window.

3.7.3

Method window - Live modifications

Subwindow: Workplace

Method

Double-clicking on a command in the method window or in the context-
sensitive menu item Properties will open the properties window of the
relevant command in which all of the parameters are shown, but in which
only the specified live parameters can be edited. In addition, this parameters
window contains the [Apply] button with which the modified live param-
eters can also be applied when the parameters window is open. Live
parameters can be modified not only in the BUSY status but also in the
READY status. If the option Comment on modification of methods is
activated in the Security settings, then the dialog window Modification
comment for method
is displayed before the modified data is applied.

The modified parameters apply until either the method is changed, the user
signs off or the program is closed. A message appears in such cases asking
whether the method should be resaved. Responding with [Yes] causes a
new method version to be created; responding with [No] retains the old
method version.

If live parameters are modified, then these modifications will be documented
not only in the determination but also in the audit trail. All parameters which
were modified will be marked with an asterisk (*) in the parameters report
for the determination and saved at the end of the determination. If param-
eters of already processed commands are modified, then these changes will
not in fact take effect until the next determination, although they will nev-
ertheless be marked as "modified live" in the method report for the deter-
mination that has been executed. The markings disappear as soon as the
method is saved in the regular way.

Methods with modified live parameters can also be resaved manually at any
time in the READY status withFile, Method, Save orFile, Method, Save
under…
.

Note

No live parameters can be edited with methods which have been signed
at Level 2 and thus locked.

Errors whose content is defined by a formula cannot be modified live.

If an active SEQUENCE command is canceled, then only the individual
command in the sequence that is currently active will be canceled, and
the next individual command will be started.