5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

449

5.6.3.2.2.7

DET pH - Additional evaluations

Tab: Method

DET pH Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab:

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value
, Volume or Time) for the fix endpoint from the measuring
point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.2.2.11, page 456).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.2.2.11, page 456).

[Delete]

Deletes the selected line.

pK/HNP evaluation

pK/HNP evaluation

on | off (Default value: off)

If this option is enabled, then the pK value is determined from the titration
curve which corresponds to the pH value at the half neutralization point
(see Chapter 5.6.3.9.1, page 854).