5 properties - history, History – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1308

Advertising
background image

6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1292

■■■■■■■■

tiamo 2.1 (ProcessLab)

message appears and you can choose whether you will nevertheless accept
the value or whether the previous value is to be retained.

Lower limit

Lower limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

-2000.0 mV

Upper limit

Upper limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

2000.0 mV

Cell constant monitoring (only for conductivity sensors)
on | off (Default value: off)

If this option is switched on, then a check will be made to see whether the
value determined during the determination of the cell constant of a con-
ductivity sensor is within the defined limits. If it is outside them, a corre-
sponding message appears and you can choose whether you will
nevertheless accept the value or whether the previous value is to be retained.

Lower limit

Lower limit for cell constant.

Range

0.001 ... 500 /cm

Default value

0.001 /cm

Upper limit

Upper limit for cell constant.

Range

0.001 ... 500 /cm

Default value

0.001 /cm

6.7.3.5

Properties - History

Tab: Configuration

Sensors Edit Sensor History

Note

This tab only appears for pH and ISE electrodes.

History table
The table contains the last 10 calibrations for the selected sensor and cannot
be edited or sorted. The individual calibrations are arranged chronologically
according to date so that the most recent determination appears last. For

Advertising