background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1071

Titrando, 855
Range

0.1 ... 999999 s

Default value

5.0 s

Titrino
Range

1 ... 999999 s

Default value

5 s

5.6.6.3.5.4

DOS U - Stop conditions

Tab: Method

DOS U Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the dosing are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the dosing.

Stop volume

Stops when the given volume has been added after the start of the dosing.
The stop volume should be adapted to suit the measuring vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100.000 mL

Selection

off

Titrino
Range

0 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Range

0 ... 999999 s

Selection

off

Default value

off

off
Stops when the preset time has elapsed since the start of the dosing.

Filling rate

Rate at which the dosing cylinder is to be filled after dosing. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used. If the entered filling rate is too high for the selected dosing