5 prep, 1 prep - overview, Prep – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1098

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1082

■■■■■■■■

tiamo 2.1 (ProcessLab)

Dosing

Range

0.00000 ... 99999.9 mL

Default value

10.0000 mL

Aspirate

Range

0.00000 ... 50.0 mL

Default value

10.0000 mL

Rate

Speed at which the commands Dosing, Fill, Aspirate, Eject and
Exchange position are carried out. The maximum rate depends on the
cylinder volume of the exchange unit or dosing device used. Aspiration and
Ejection of the sample should be done at a rate <10 mL/min.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

5.6.6.5

PREP

5.6.6.5.1

PREP - Overview

Dialog window: Method

PREP Properties... PREP - 'Command name'

Command for Rinsing cylinder and tubing of an exchange or dosing unit.

Devices
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 841, 842, 851, 852, 857, 888, 890, 901,
902, 904, 905, 906, 907

Titrino: 736, 751, 758, 799

Dosing Interface: 846

Conductometer: 856

pH/Ion - Meter: 867

Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Advertising