3 meas t/flow - measuring parameters, 4 meas t/flow - evaluations, Measuring parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 943: Evaluation

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

927

5.6.4.7.3

MEAS T/Flow - Measuring parameters

Tab: Method

MEAS T/Flow Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measuring time

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

600 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

1 ... 999999.0 s

Default value

10 s

Stop measured value Temp.

Stops when the preset measured value has been reached since the start of
the measurement. off means no stop.

Range

50.0 ... 250.0 °C

Selection

off

Default value

off

5.6.4.7.4

MEAS T/Flow - Evaluations

Tab: Method

MEAS T/Flow Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves can
be activated and defined on this tab.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the minimum measured
value will be interpolated from the measuring point list.

Advertising