5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

622

■■■■■■■■

tiamo 2.1 (ProcessLab)

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos
and 855.

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command. This parameter is displayed only for 751, 785, 798 and 799.

5.6.3.3.5.3

MET Upol - Start conditions

Tab: Method

MET Upol Properties... Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the titration
is started.

Initial measured value

Note

Is displayed only for Titrando and 855.

The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters: