5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

792

■■■■■■■■

tiamo 2.1 (ProcessLab)

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

15 s

Pause

Pause

Waiting time before the start of the titration.

Range

0 ... 999999 s

Default value

0 s

5.6.3.7.4

BRC - Control parameters

Tab: Method

BRC Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Endpoint

Ep with

Measured value for the endpoint.

Range

-1250.0 ... 1250.0 mV

Default value

200.0 mV

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.