3 stat u - start conditions, Start conditions – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 853

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

837

Titrando, 855
Selection

1 | 2 | 3 | 4 | off

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos
and 855.

Switch on/off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the end
of the command. This parameter is displayed only for 751 and 799.

5.6.3.8.7.3

STAT U - Start conditions

Tab: Method

STAT U Properties... Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the titration
is started.

Initial measured value

Note

Is displayed only for Titrando and 855.

The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:

Advertising