8.4 Manual Control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1494

■■■■■■■■

tiamo 2.1 (ProcessLab)

8.4.6.3.2

Digital output

Program part: Manual Control

If in the device tree a single digital output is selected, then the properties of
the digital output on the tabs Switch on/off, Timer control and Set
default
are displayed in the subwindow for functions/parameters.

In the subwindow for the graphical display, the status of the input is dis-
played.

Switch on/off

Port description

Display of the port description.

Port

Display of the internal description of the port.

Type

Display of the port type.

Parameter 1

The meaning of Parameter 1 depends on the Type of the digital output that
has been set.

When Type = Stirrer: stirrer output in %, pulse duration dependent on
Parameter 2.

Range

-214783647 ... 214783647

Default value

100

Parameter 2

The meaning of Parameter 2 depends on the Type of the digital output that
has been set.

For Type = Stirrer:

0: Duration of the switch-on pulse fixed at 500 ms.

>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in 10
ms
Example:
Parameter 1
= 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on 225
ms, switched off 75 ms

Range

-214783647 ... 214783647

Default value

0