5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

858

■■■■■■■■

tiamo 2.1 (ProcessLab)

5.6.3.10

Additional measured values

5.6.3.10.1

Calculated measured value

Dialog window: Method

DET/MET/SET/KFT/KFC/STAT/MEAS/DOS Proper-

ties...

Additional measured values [New]/[Properties] Calculated measured

value

With the option Additional calculated measured values, a maximum of
3 new measured values can be calculated from existing measured values and
saved, which can also be presented as curves under the designation Calcu-
lated 1…3
and used in formulasas variables *.CA1…3.

Name

Designation for the calculated measured value. This name is used as axis
legend.

Input

25 characters

Default value

Calculated measured value #

Formula

Formula to calculate the measured value. It can either be entered directly or

can be created with the formula editor which opens by pressing

.

Input

1000 characters

Default value

'empty'

Unit

Unit of calculated measured value.