12 meas spec, 1 meas spec - overview, Meas spec – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 975

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

959

End wavelength

Upper limit of the spectrum.

Range

200.0 ... 1100.0 nm

Default value

1000.0 nm

Integration time

Integration time for spectrum recording.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Range

0 ... 100 Pixels

Default value

0 Pixels

5.6.4.12

MEAS Spec

5.6.4.12.1

MEAS Spec - Overview

Dialog window: Method

MEAS Spec Properties... MEAS Spec - 'Command

name'

Command for the Measurement of the sample spectrum.

Devices
This command can be executed with the following device:

Spectrometer: Avantes

Appearance
The command has the following appearance:

Parameters
The parameters for the command MEAS Spec are configured on the fol-
lowing 3 tabs:

Advertising