5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1108

■■■■■■■■

tiamo 2.1 (ProcessLab)

Off causes heating at the maximum possible heating rate (see device man-
ual).

Range

1 ... 999 min

Selection

off

Default value

off

Timeout

If the Target temperature is not reached within the defined Heating
time
then this waiting time will start. This parameter is only visible if a
Heating time has been defined.

Range

1 ... 999 min

Selection

off

Default value

off

Action if timeout expires

on | off (Default value: on)

If this option is enabled, then an action can be defined which is to be carried
out if the target temperature has not yet been reached after expiry of the
maximum waiting time (is displayed only when Timeoutoff).

Selection

Display message | Stop determination | Stop
determination and series

Default value

Display message

Display message
A message is displayed and you can choose whether you want to con-
tinue the run or cancel it.
Stop determination
The determination is stopped and the next determination of the series is
started.
Stop determination and series
Both the determination and the series are stopped and the exit track will
be started (if available).

5.6.7.9

FLOW

5.6.7.9.1

FLOW - Overview

Dialog window: Method

FLOW Properties... FLOW - 'Command name'

Command for controlling the gas flow.

Devices
This command can be executed with the following devices:

Sample Processor: 774, 874